- Overview
- Intraoral scanners in Australia
- How to check a scanner or other dental device on the ARTG
How to check a scanner or other dental device on the ARTG
In short
An ARTG entry shows that a kind of medical device is on the register for one sponsor. Under the Therapeutic Goods Act, devices are the same kind when they share the sponsor, the manufacturer, the nomenclature code and the class, so one entry can cover several models. For Class IIa, IIb and III devices the code is a Global Medical Device Nomenclature preferred term; for Class I devices it is a template term, unless the device is supplied sterile, has a measuring function or has no template term. Only Class III devices, and some diagnostic devices, add the unique product identifier. Our advice: ask the seller for the ARTG number that covers your exact model, check the sponsor and manufacturer on the entry, and get the seller to confirm the model in writing.
In this article
Before you buy an intraoral scanner, an X-ray unit or most other clinical equipment in Australia, it is worth checking the Australian Register of Therapeutic Goods (ARTG). The check takes a few minutes. Reading the entry correctly is the harder part, because an entry usually covers a kind of device, not one model.
Our position: an ARTG entry under the right sponsor is a starting point, not proof that the model on your quote is covered. Ask the seller for the ARTG number for your exact model and get that answer in writing.
What the ARTG is
Section 9A of the Therapeutic Goods Act 1989 sets up the ARTG: the Secretary of the federal health department “is to cause to be maintained a register, to be known as the Australian Register of Therapeutic Goods”. Section 9E requires a list of the goods on the register to be published at least once every 12 months. In practice, you search it on the website of the Therapeutic Goods Administration (TGA). Australian Register of Therapeutic Goods (ARTG) | Therapeutic Goods Administration (TGA) (Therapeutic Goods Administration (TGA))1
Who answers for what: sponsor, manufacturer and seller
Three companies can stand behind one scanner, and they can all be different.
The sponsor is defined in section 3 of the Therapeutic Goods Act 1989 as including the person who imports the goods into Australia or arranges their import, and the person who manufactures them in Australia, or arranges their manufacture, for supply. A company that imports on behalf of another company that is resident or carries on business in Australia is not the sponsor. The ARTG entry names the sponsor.
The manufacturer, under section 41BG of the same Act, is the person “responsible for the design, production, packaging and labelling of the device before it is supplied under the person’s name”, whether it does that work itself or has someone else do it. It can be overseas.
The seller you buy from may be neither. Ask it what it covers itself: the warranty, the service and any loan unit.
So when a dealer says a scanner is “TGA approved”, ask which company is the sponsor on the entry, who services the scanner in your state, and who you contact if there is a recall. If the three answers name three companies, that is normal, but you want the names in writing.
Why one entry can cover many models
The Therapeutic Goods Act 1989, section 41BE, says one medical device is “of the same kind” as another if they have the same sponsor, the same manufacturer, the same device nomenclature system code and the same medical device classification, plus any other characteristics the Regulations prescribe.
The Therapeutic Goods (Medical Devices) Regulations 2002 fill in the detail:
- Regulation 1.7 takes the code from the Global Medical Device Nomenclature. For Class IIa, IIb and III devices the code is the “preferred term”. Class I devices use the “template term”, unless they are supplied sterile, have a measuring function or have no template term; those use the preferred term too.
- Regulation 1.6 adds the unique product identifier as a characteristic only for Class III devices and some diagnostic (IVD) devices.
So for a Class I, IIa or IIb device, every model with the same sponsor, manufacturer, code and class can sit under one entry. The entry may not name your model at all. That is how the law is meant to work, but it means the register on its own cannot tell you which models a seller has put under an entry.
Five checks before you sign
- Get the number. Ask the seller for the ARTG number that covers the exact model and version on the quote.
- Open the entry. Search the number on the TGA site and read what the entry lists.
- Match the sponsor. Check that the sponsor on the entry is the company the seller named. If the seller is not the sponsor, ask who is, and who handles recalls and complaints.
- Match the manufacturer and the kind. Check that the manufacturer on the entry is the company named as manufacturer on the device label. The kind of device and the intended purpose should describe what you are buying.
- Get the model in writing. If the entry doesn’t name your model, ask the seller, or the sponsor, to confirm in writing that the entry covers it. Keep that with the invoice.
If a number you were given does not show up, don’t assume the worst straight away. The entry may have been cancelled, for example because the sponsor asked for it or stopped paying the annual charge (section 41GL). Section 41GP of the Therapeutic Goods Act 1989 requires the Secretary to publish a notice of each cancelled medical device entry “in the Gazette or on the Department’s website”. Ask the seller why the number is missing, and ask the TGA if the answer doesn’t add up.
Why it is worth the five minutes
Section 41MI of the Therapeutic Goods Act 1989 makes it an offence to import or supply a medical device in Australia that is not of a kind included in the Register for that person, unless an exemption, approval or authority applies. Under section 41MI(6), a person who proves it was not the sponsor of the device has a defence. That does not help your practice if a device it relies on has to be withdrawn.
A seller that sells a device legally can answer the five checks in one email. If it can’t name an ARTG number for the model it is quoting, choose another seller or another model.
See the comparison with sources and dates: Intraoral scanners in Australia.
Related articles
- Intraoral scanner prices in Australia: what sellers actually publish
Which intraoral scanner prices Australian sellers publish, what they leave out, the warranty terms on paper, and what to ask for in a quote.
Sources
Numbered by first appearance on the page. The date is when the source was last checked.
- Australian Register of Therapeutic Goods (ARTG) | Therapeutic Goods Administration (TGA), Therapeutic Goods Administration (TGA). Checked 24 October 2026. Official sourceConfirmed in the source