Skip to content

Dental X-ray shielding requirements in Australia: when you need a shielding assessment

By Ole Tom Pettersen, Dental Markedsføring · published · page last reviewed

In short

In Victoria, a management licence application for extra-oral X-ray equipment, such as an OPG or CBCT, must include a shielding assessment report, and the Department of Health published a new shielding assessment standard on 24 September 2026. In Western Australia, a third-party radiation consultant must assess the room for a CBCT, and the Radiological Council reviews it before the unit can be approved for use. NSW EPA Radiation Guideline 7 sets minimum shielding assessment requirements for premises where radiation apparatus is to be installed. Our advice: get the shielding assessment done on the floor plan before the builder starts, and agree in writing who orders and pays for it.

A shielding assessment works out how much shielding the walls, doors, floor and ceiling of an X-ray room need. For an OPG or a CBCT it is often a document the regulator wants to see before you can use the unit. It also tells the builder what to build, so the order of steps matters.

Our position: commission the shielding assessment on the floor plan, before the fit-out is priced. Shielding is built into the walls and doors, so it is easier to plan in than to add later, and in Victoria and Western Australia the regulator reviews the assessment before you can use the unit.

Victoria: a report with every extra-oral licence application

The Department of Health’s page for dental practices says that “If you are applying for a licence involving extra-oral radiation apparatus you must submit a shielding assessment report with your application.” It applies to extra-oral X-ray apparatus only, so an intra-oral unit on its own does not need one under this rule.

The report must include:

  1. A floor plan of the X-ray room and the rooms around it.
  2. Occupancy factors and the purpose of the surrounding rooms.
  3. The method and references used.
  4. The shielding required for the operator.
  5. The workloads assumed for the equipment.
  6. Where related equipment sits, such as controls, protective screens and warning lights.

The department published a Radiation Shielding Assessment Standard on 24 September 2026. It sets “the minimum standard for conducting and reporting shielding assessments in Victoria”, and says licence conditions may also require shielding assessments to be “maintained, implemented and kept current”. If your plans were drawn before that date, ask your assessor whether the report meets the new standard.

Western Australia: a consultant’s assessment before CBCT approval

The Radiological Council’s specific requirements page says a radiation shielding assessment “will need to be conducted by a third-party radiation consultant for the room where the CBCT equipment will be installed”, and that it “must be submitted to the Radiological Council for review before a CBCT unit can be approved for use”. The same page says evidence of a reporting agreement with a radiologist must be submitted with the application to register CBCT equipment.

New South Wales: the EPA shielding guideline

The NSW EPA’s Radiation Guideline 7 sets “the minimum shielding assessment requirements for premises where” radiation apparatus, including diagnostic imaging apparatus, is to be installed. The EPA says the guideline does not apply to existing premises where shielding is already in place, “unless there are changes to the existing building, equipment or the use of surrounding areas”. So replacing an OPG with a CBCT, or turning a storeroom next to the X-ray room into a surgery, can bring it back into play. The page was last updated on 14 July 2021 and says the guideline is intended to become mandatory for new premises, so ask the EPA or your consultant how it applies to your project.

Where it fits in the fit-out

  1. Choose the unit and the room, and get the floor plan drawn with the rooms around it labelled.
  2. Commission the shielding assessment from a radiation consultant, using that plan and the unit’s specifications.
  3. Give the report to the builder before the walls and doors are priced.
  4. Submit the report with your licence or registration application where your state requires it.
  5. Keep the report with your radiation records, and have it reviewed if you change the unit or the use of the rooms around it.

Before you sign for a new OPG or CBCT, ask the seller in writing whether it arranges the shielding assessment or leaves it to you, and who pays for it. None of the four regulator pages we read on 7 November 2026 gives a price for a shielding assessment, so get a quote from the consultant before you budget.